Model Number UM-S20-17S
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported, the ultrasonic probe outer covering on insertion tube was cracked.The issue did not occur during a procedure.There were no reports of patient harm.
 
Manufacturer Narrative
Three attempts were performed to obtain additional information, but no response was received from the customer.The device was returned to olympus for evaluation and the customer¿s allegation was confirmed.The device evaluation found the outer covering of the insertion tube was cracked.No further testing could be performed due to the broken insertion tube.Based on the results of the investigation, it is likely that the following led to the malfunction: component failure.A definitive root cause could not be identified.A device history review revealed no issues that could have caused or contributed to the reported issue.Should any additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19100914
MDR Text Key340804584
Report Number3002808148-2024-32181
Device Sequence Number1
Product Code ITX
UDI-Device Identifier04953170368479
UDI-Public04953170368479
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K982323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2024
Initial Date Manufacturer Received 03/15/2024
Initial Date FDA Received04/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-