Model Number 7841
Device Problems Failure to Capture (1081); Device Sensing Problem (2917); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2024
Event Type  Injury  
Event Description
It was reported that this patient presented to the clinic due to a right atrial (ra) lead repositioning procedure.It was noted that there was no p wave sensing and no capture from the pre procedure check and fluoroscopy confirmed that the lead had dialoged from its original implant location.The lead was successfully repositioned to the lateral atrial wall and all measurements were appropriate.No additional adverse patient effects were reported.
 
Event Description
It was reported that this patient presented to the clinic due to a right atrial (ra) lead repositioning procedure.It was noted that there was no p wave sensing and no capture from the pre procedure check, and fluoroscopy confirmed that the lead had dislodged from its original implant location.The lead was successfully repositioned to the lateral atrial wall and all measurements were appropriate.No additional adverse patient effects were reported.
 
Manufacturer Narrative
This report is being filed to update the implant date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGEVITY+
Type of Device
DRUG ELUTING PERMANENT RIGHT VENTRICULAR OR ATRIAL PACEMAKER ELECTRODES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19102185
MDR Text Key340100133
Report Number2124215-2024-22426
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number7841
Device Catalogue Number7841
Device Lot Number1371179
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/11/2024
Initial Date FDA Received04/12/2024
Supplement Dates Manufacturer Received04/11/2024
Supplement Dates FDA Received04/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/16/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-