ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; Implantable pulse generator, pacemaker (non-CRT) Back to Search Results
Model Number PM2272
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/25/2024
Event Type  Injury  
Event Description
Related manufacturer reference number: 2017865-2024-38331 it was reported the device is subject to the assurity and endurity pacemakers header anomaly advisory issued by abbott on 15 march 2021.The pacemaker was explanted prophylactically with no known malfunctions.Device interrogation revealed noise oversensing and low pacing impedance on the right ventricular (rv) lead.On (b)(6) 2024 the physician elected to cap and replace the rv lead.The patient was in stable condition.
 
Manufacturer Narrative
The device was returned due to advisory with no allegation of a malfunction.Visual inspection of the device and header attachment area did not find any anomalies.Electrical, longevity, and visual analysis was normal with no anomalies found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
Implantable pulse generator, pacemaker (non-CRT)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19102298
MDR Text Key340100959
Report Number2017865-2024-38330
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734509589
UDI-Public(01)05414734509589(10)A000093219(17)210831
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2021
Device Model NumberPM2272
Device Lot NumberA000093219
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/25/2024
Initial Date FDA Received04/12/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2088TC/46
Patient Outcome(s) Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-