C.R. BARD, INC. (BASD) -3006260740 POWER PORT MRI ISP, 8 FR GROSHONG, INT WO SP, SL; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 8808561 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/31/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiry date: 10/2023).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
|
Event Description
|
It was reported that prior to a port placement procedure via the right internal jugular vein, while the surgeon was flushing each parts of the package, black substances were allegedly found to be flushed out of the catheter during flushing.The procedure was completed using another device.There was no patient contact.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one power port isp mri implantable port with groshong catheter with several components were returned for sample evaluation.Visual and functional evaluation were performed.The samples appeared to be clean.An attempt to flush the groshong catheter, sheath and vessel dilator with water were performed.No anomalies were observed.Therefore, the investigation is unconfirmed for reported black substance flushed out issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 10/2023).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that prior to a port placement procedure via the right internal jugular vein, while the surgeon was flushing each part of the package, black substances were allegedly found to be flushed out of the catheter during flushing.The procedure was completed using another device.There was no patient contact.
|
|