• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE AUTOGUARD WINGED; PERIPHERAL IV CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE AUTOGUARD WINGED; PERIPHERAL IV CATHETERS Back to Search Results
Catalog Number 381912
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Manufacturer Narrative
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Event Description
It was reported that bd insyte autoguard winged unable to connect cap to adapter.The following information was provided by the initial reporter, translated from chinese to english: on (b)(6) 2024, the nurse followed the doctor's orders for the child to draw blood for laboratory tests, using a straight indwelling needle to puncture, after successful puncture, ready to connect the heparin cap, found that there is a defect in the connecting port of the indwelling needle, and the heparin cap can not be connected, the incident immediately after the use of 2.5ml syringe connected to the indwelling needle to draw blood, and after the completion of the drawing of blood, the indwelling needle was withdrawn.No impact on children's health.
 
Manufacturer Narrative
Investigation results: no photographic evidence or physical samples were accessible to investigate the reported condition.Our quality engineer team conducted a comprehensive review of the device history record for the provided material number 381912 and lot number 3248257.This review did not uncover any abnormalities during the production process that could have caused this defect, and all quality tests were found to be within the specified standards.Unfortunately, the exact cause of this incident could not be determined based on the information available.We will continue to track and analyze complaints related to this device and the reported condition in order to identify any emerging trends.
 
Event Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE AUTOGUARD WINGED
Type of Device
PERIPHERAL IV CATHETERS
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key19102699
MDR Text Key340811456
Report Number1710034-2024-00331
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00382903819126
UDI-Public(01)00382903819126
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
UNK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number381912
Device Lot Number3248257
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/25/2024
Initial Date FDA Received04/13/2024
Supplement Dates Manufacturer Received05/07/2024
Supplement Dates FDA Received05/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/07/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-