Philips received a complaint on the defibrillator indicating ecg failed to show battery test.Information regarding specified malfunction has been requested.There was no patient involvement.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
|
Analysis was performed by onsite service personnel, followed the troubleshooting processes and flowcharts indicated in the service manual: executed the performance verification procedures indicated, so visual inspection (device, cables, supplies and accessories), all service mode tests (operational check, printer test, controls test, display test, fan test and usb test), functional checks (all parameters check, defibrillator measurement test, defibrillator test with ac and battery, defibrillator charge cancellation test, pacer test, synchronized cardioversion test and paddles safety check) and used previous field experience.Based on the results of the analysis, it was determined that the product has malfunctioned and the cause of the reported problem was battery can't charge.The issue was resolved by replaced 989803190371 dfm100 battery.
|