• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC. EFFICIA DFM100; XL+DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC. EFFICIA DFM100; XL+DEFIBRILLATOR Back to Search Results
Model Number 866199
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2024
Event Type  malfunction  
Event Description
Philips received a complaint on the defibrillator indicating ecg failed to show battery test.Information regarding specified malfunction has been requested.There was no patient involvement.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
 
Manufacturer Narrative
Reporting address line 1: (b)(6).Reporter city: (b)(6).Reporting address state: (b)(6).Reporting address postal: (b)(6).Reporting institution phone: (b)(6).Reporter phone: (b)(6).A follow up report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
Analysis was performed by onsite service personnel, followed the troubleshooting processes and flowcharts indicated in the service manual: executed the performance verification procedures indicated, so visual inspection (device, cables, supplies and accessories), all service mode tests (operational check, printer test, controls test, display test, fan test and usb test), functional checks (all parameters check, defibrillator measurement test, defibrillator test with ac and battery, defibrillator charge cancellation test, pacer test, synchronized cardioversion test and paddles safety check) and used previous field experience.Based on the results of the analysis, it was determined that the product has malfunctioned and the cause of the reported problem was battery can't charge.The issue was resolved by replaced 989803190371 dfm100 battery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EFFICIA DFM100
Type of Device
XL+DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen
CH 
Manufacturer (Section G)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC.
no.2 keji north 3rd road
nanshan district
shenzhen
CH  
Manufacturer Contact
feng she
no.2 keji north 3rd road
nanshan district
shenzhen 
CH  
7552698099
MDR Report Key19103300
MDR Text Key340908392
Report Number3030677-2024-01368
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866199
Device Catalogue Number866199
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/18/2024
Initial Date FDA Received04/15/2024
Supplement Dates Manufacturer Received05/22/2024
Supplement Dates FDA Received05/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-