CORIN LTD TRINITY; TRINITY-I ACETABULAR SHELL Back to Search Results
Model Number 321.02.346
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/02/2024
Event Type  malfunction  
Event Description
During surgery once the trinity acetabular shell was impacted in the acetabulum, it could not be detached from the trinity introducer handle.The impacted shell was removed and alternatives used.This caused a delay to surgery of more than 30 minutes.
 
Manufacturer Narrative
(b)(4) initial report.The reported devices are being returned to corin and will be examined.Details of this review will be provided in a supplemental report upon completion of the investigation.The appropriate device details have been provided and the relevant device manufacturing records will be identified and reviewed.Please note: this report is filed with the fda due to an event experienced with a device that is similar to those placed on the market in the usa, however, this event occurred outside of the usa.The submission of this report does not constitute an admission that the device, reporting entity, entity's representative or distributor caused or contributed to this event.
 
Manufacturer Narrative
Per -5768 final report the reported trinity introducer handle and trinity acetabular shell were returned for examination.The failure mode was confirmed, however, after a few impactions the shell could be removed from the handle.The appropriate device details were provided and the relevant device manufacturing records have been identified and reviewed.All parts associated with these records conformed to material and dimensional specification at the time of manufacture.Based on the available information, the root cause of the event could not be determined.However, trending has identified that this could be a user issue.Corin will continue to monitor trends relating to this or similar failure modes and this case is now considered closed.Please note: this report is filed with the fda due to an adverse event experienced with a device that is similar to those placed on the market in the usa, however, this event occurred outside of the usa.The submission of this report does not constitute an admission that the device, reporting entity, entity's representative or distributor caused or contributed to this event.
 
Event Description
During surgery once the trinity acetabular shell was impacted in the acetabulum, it could not be detached from the trinity introducer handle.The impacted shell was removed and alternatives used.This caused a delay to surgery of more than 30 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRINITY
Type of Device
TRINITY-I ACETABULAR SHELL
Manufacturer (Section D)
CORIN LTD
the corinium centre
cirencester, gloucestershire GL7 1 YJ
UK  GL7 1YJ
Manufacturer (Section G)
CORIN LTD
the corinium centre
cirencester, gloucestershire GL7 1 YJ
UK   GL7 1YJ
Manufacturer Contact
dardan uka
the corinium centre
cirencester, gloucestershire GL7 1-YJ
UK   GL7 1YJ
MDR Report Key19103408
MDR Text Key340817890
Report Number9614209-2024-00315
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K122305
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number321.02.346
Device Catalogue NumberNOT APPLICABLE
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/02/2024
Initial Date FDA Received04/15/2024
Was Device Evaluated by Manufacturer? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TRINITY INTRODUCER HANDLE: 9210129, 095751-05; TRINITY INTRODUCER HANDLE: 9210129, 095751-05
Patient Outcome(s) Required Intervention;
-
-