• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI CAPSURE EPICARDIAL LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI CAPSURE EPICARDIAL LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 4968-35
Device Problems High impedance (1291); Capturing Problem (2891); Impedance Problem (2950)
Patient Problems Bradycardia (1751); Dizziness (2194)
Event Date 04/02/2024
Event Type  Injury  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that the patient experienced a low heart rate and dizziness.An interrogation report indicated that the right ventricular (rv) exhibited an acute rise in thresholds and  impedance and triggered an alert for high impedance.The lead remains in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was further reported that the patient presented to the emergency department and was admitted.The rv lead was reprogrammed from bipolar to unipolar, which improved capture and sensing.The rv lead remains in use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE EPICARDIAL LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key19103934
MDR Text Key340490418
Report Number2649622-2024-10276
Device Sequence Number1
Product Code DTB
UDI-Device Identifier00613994492302
UDI-Public00613994492302
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P950024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/06/2018
Device Model Number4968-35
Device Catalogue Number4968-35
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/02/2024
Initial Date FDA Received04/15/2024
Supplement Dates Manufacturer Received04/19/2024
Supplement Dates FDA Received05/08/2024
Date Device Manufactured11/02/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
(B)(6) IPG
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age14 YR
Patient SexMale
-
-