MAUDE Adverse Event Report: MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Catalog Number CLXECP
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2024
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction centrifuge bowl leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot m128 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot m128 shows no trends.Trends were reviewed for complaint category, centrifuge bowl leak/break.No trends were detected for this complaint category.Photographs were provided by the customer for evaluation.A visual inspection of the customer-supplied photographs showed that the drive tube and centrifuge bowl are damaged.The centrifuge bowl had dislodged out of the bowl holder and impacted the inside of the centrifuge chamber during the treatment.A known cause for the centrifuge bowl to dislodge out of the bowl holder is due to the bowl not being properly locked into the bowl holder during installation by the end user.A material trace of the bowl assembly and its components used to build lot m128 found no related non-conformances.The device history record (dhr) review did not identify any related non-conformances, deviations, or equipment maintenance events.This lot passed all lot release testing.The most likely root cause of the centrifuge bowl break is due to the centrifuge bowl not being secured into the bowl holder during installation of the kit by the end user.No further action is required at this time.The investigation is now complete.(b)(4).(b)(4) 15-apr-2024.
 
Event Description
The customer contacted mallinckrodt to report they experienced a centrifuge bowl leak/break with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported they observed a bowl break at the beginning of the treatment.The ecp treatment was aborted and residual blood within the kit was not returned to the patient.The patient was reported to be in stable condition.The kit return was requested; however, the customer discarded the kit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
blanchardstown, dublin, D15 T X2V
EI  D15 TX2V
Manufacturer (Section G)
HARMAC MEDICAL PRODUCTS
i.d.a. business park
castlerea roscommon,
EI  
Manufacturer Contact
mark wendelken
440 route 22 east
suite 302
bridgewater, NJ 08807
MDR Report Key19110673
MDR Text Key340812198
Report Number3013428851-2024-00029
Device Sequence Number1
Product Code LNR
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCLXECP
Device Lot NumberM128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/28/2024
Initial Date FDA Received04/15/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
Patient Weight40 KG
-
-