• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS (TPE); SEPARATOR FOR THERAPEUTIC PURPOSES, MEMBRANE AUTOMATED BLOOD CELL/PLASMA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS (TPE); SEPARATOR FOR THERAPEUTIC PURPOSES, MEMBRANE AUTOMATED BLOOD CELL/PLASMA Back to Search Results
Catalog Number 107144
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/21/2024
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an external fluid leak was observed from the dialysate port and housing connection point of a prismaflex tpe2000 set during priming.S the filter into the return line.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H11: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional air leak test (2 bar pressure) was performed and a leak was observed at the level of the connection from the filter to the effluent port due to a crack.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the damage was determined to be related to mechanical shock applied to the product during shipping, transport, and/or improper storage.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SETS (TPE)
Type of Device
SEPARATOR FOR THERAPEUTIC PURPOSES, MEMBRANE AUTOMATED BLOOD CELL/PLASMA
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19110759
MDR Text Key340240810
Report Number8010182-2024-00170
Device Sequence Number1
Product Code MDP
UDI-Device Identifier07332414069339
UDI-Public(01)07332414069339
Combination Product (y/n)N
Reporter Country CodeVM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number107144
Device Lot Number23E0045A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/28/2024
Initial Date FDA Received04/15/2024
Supplement Dates Manufacturer Received05/10/2024
Supplement Dates FDA Received05/22/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-