H11: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional air leak test (2 bar pressure) was performed and a leak was observed at the level of the connection from the filter to the effluent port due to a crack.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the damage was determined to be related to mechanical shock applied to the product during shipping, transport, and/or improper storage.Should additional relevant information become available, a supplemental report will be submitted.
|