DEPUY ORTHOPAEDICS INC US GLOBAL ADV HUM HEAD DISTRACTOR; EXTRACTION INSTRUMENTS Back to Search Results
Catalog Number 212801021
Device Problems Break (1069); Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2024
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
This instrument is composed of 1 part.The tip of the instrument is chipped and dull.The product damage documented in this pc has been identified during loan kit inspection, by the loan kit technician.The event date and alert date are the date that the inspection took place.There is no surgeon, procedure, or patient details available.No further information can be obtained as the case was not reported by the customer.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, this instrument is composed of 1 part.The tip of the instrument is chipped and not sharp as per the images attached.The product was not returned to depuy synthes, however photos were provided for review.See attachment (image (1).Jpg and image (2).Jpg).The photo investigation revealed that global adv hum head distractor had broken chipped the tip in addition, wear is observed on the tip.Based on the observed condition, it is possible to confirm that the cutting edge is dull.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.Since the device was not returned, a dimensional inspection cannot be performed.A functional test was unable to be performed due to the device was not returned.The overall complaint was confirmed as the observed condition of the global adv hum head distractor would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLOBAL ADV HUM HEAD DISTRACTOR
Type of Device
EXTRACTION INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key19112449
MDR Text Key340481166
Report Number1818910-2024-08342
Device Sequence Number1
Product Code HWB
UDI-Device Identifier10603295088035
UDI-Public10603295088035
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K170748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number212801021
Device Lot NumberAK1105
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/11/2024
Initial Date FDA Received04/15/2024
Supplement Dates Manufacturer Received05/09/2024
Not provided
Supplement Dates FDA Received05/09/2024
05/29/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-