MEDTRONIC SINGAPORE OPERATIONS AZURE XT DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE
|
Back to Search Results |
|
Model Number W2DR01 |
Device Problem
Migration or Expulsion of Device (1395)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/10/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
It was reported that five days post implant during a pre-discharge check, the right ventricular (rv) lead exhibited rising thresholds.A chest x-ray showed the implantable pulse generator (ipg) pocket was lowered significantly to the lower part and as a result, it seemed like the entire lead was pulled.The rv lead and right atrial (ra) lead were dislodged confirmed on x-ray.The ipg and the leads were repositioned.No patient complications have been reported as a result of this event.
|
|
Search Alerts/Recalls
|
|
|