C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP MRI 8CF INT. W/SP, ATT, SL; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 8808060 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/28/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
As the lot number for the device was provided, a review of the device history records will be performed.The return of the sample is pending.The investigation of the reported event is currently underway.D4 (expiration date: 03/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a port placement procedure, the catheter lock allegedly did not feel secure when inserted over catheter end.The procedure was completed by using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: two cath-locks inside of a specimen cup were received for evaluation.Visual evaluation was performed.The complaint samples appeared to have residue throughout.No other anomalies were noted.Therefore, the investigation is inconclusive for the reported cathlock loose connection.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a port placement procedure, the catheter lock allegedly did not feel secure when inserted over catheter end.The procedure was completed by using another device.There was no reported patient injury.
|
|