Catalog Number 254500546 |
Device Problem
No Apparent Adverse Event (3189)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Event Description
|
The following devices were received.Product code: 254500548, lot no - mvmbsv780 product code: 227430500, lot no - pg308536 product code: 254500546, lot no - mvmbwq420 product code: 254500548, lot no - mvmbxb160 product code: 254500548, lot no - mvmhmc720 however, they couldn't be associated with a complaint or any other existing record.As a result, this complaint was created to analyze the returned device.This complaint involves one (5) devices.
|
|
Manufacturer Narrative
|
Product complaint (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional event information: based on the findings from the physical evaluation: ip- (b)(4) spring missing, ip (b)(4) small post was broken, and the spring was missing, ip (b)(4) spring missing, ip (b)(4) small post was cracked, wear marks as normally and the spring was missing, ip (b)(4) the shaft was broken.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: according to the information received: the following device was received as blind unit.The product was returned to depuy synthes for evaluation.Visual examination of the found that a small post was cracked and the spring was missing.No other issues were identified.The failure mode is consistent with inserting an extractor device in between the trial and mating shim and using the extractor to pry the mating devices apart during extraction of the trials.This improper technique results in damage to the spring and/or post components of the articulating surface as well as the mating shim.The attune intuition surgical technique 0612-10-512 (page 53), emphasizes the correct use of the tibial trial extractor (product code 254500138) with the trials.Furthermore, on page 51 of the surgical technique and per ifu-0902-00-836, careful inspection of the trials for damage/breakage should be performed pre and post-operative.If any damage to the balseal components is observed the trail should be replaced.In the event of instrument breakage during use, ensure that all device fragments that may have entered the surgical site are removed prior to completion of the procedure, as patient injury may result.The overall complaint was confirmed as the observed condition of the device 254500546, attune rp ps artic surf sz6 would contribute to the device issue.Based on the investigation findings, unintended use error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Search Alerts/Recalls
|
|