• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CRITICAL CARE AB FLOW-I; GAS-MACHINE, ANESTHESIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CRITICAL CARE AB FLOW-I; GAS-MACHINE, ANESTHESIA Back to Search Results
Model Number FLOW-I C20
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the sevoflurane vaporizer in the anesthesia workstation failed in system checkout and generated a technical error code indicating vaporizer error.The vaporizer is the unit containing anesthetic agent in the anesthesia workstation.When the vaporizer was received for investigation, a yellow/brown residue was found inside the vaporizer.There was no patient involvement.Manufacturer's reference #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOW-I
Type of Device
GAS-MACHINE, ANESTHESIA
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key19115236
MDR Text Key340268167
Report Number8010042-2024-00625
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K191027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFLOW-I C20
Device Catalogue Number6677200
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/02/2024
Initial Date FDA Received04/16/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-