• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLFLEX ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLFLEX ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00516900
Device Problems Positioning Failure (1158); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/20/2024
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that a wallflex esophageal partially covered stent was to be implanted in the esophagus to treat an esophageal cancer during a stent placement procedure performed on (b)(6) 2024.The patient's anatomy was not tortuous and was not dilated prior to stent placement.During the procedure, the stent was not able to be deployed, and it was noted that the white distal tip was inside the patient's esophagus under pediatric gastroscope.The guidewire was then removed, and the white distal tip came out with the wire.A different device was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Manufacturer Narrative
Block h6: imdrf device code a0501 captures the reportable event of tip detachment of device or device component.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX ESOPHAGEAL
Type of Device
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key19118250
MDR Text Key340970171
Report Number3005099803-2024-01583
Device Sequence Number1
Product Code ESW
UDI-Device Identifier08714729765240
UDI-Public08714729765240
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K073266
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00516900
Device Catalogue Number56453
Device Lot Number0032915745
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/22/2024
Initial Date FDA Received04/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/23/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-