• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_13.2
Device Problems Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Manufacturer Narrative
(b)(6).
 
Event Description
The reporter indicated that the surgeon implanted an implantable collamer lens into the patient's eye.The surgeon reports the vault is extremely high.A lens exchange was planned.Lens remains implanted.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
 
Manufacturer Narrative
B5 - the reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2; -12.50/3.0/130 (sphere/cylinder/axis) implantable collamer lens into the patient's right eye (od).The surgeon reports there is excessive vaulting with significant reduction or irido-corneal angles, shallow ac, and the patient also had ifis during surgery.On (b)(6) 2024 the lens was explanted and this resolved the problem.The cause of the event was reported as the device.H6 - health effect clinical code: 4581 - significant reduction of irido corneal angles, ifis, shallow ac.H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key19119103
MDR Text Key340818296
Report Number2023826-2024-01806
Device Sequence Number1
Product Code MTA
UDI-Device Identifier00840311312351
UDI-Public00840311312351
Combination Product (y/n)N
Reporter Country CodeSA
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVTICM5_13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/25/2024
Initial Date FDA Received04/16/2024
Supplement Dates Manufacturer Received04/28/2024
Supplement Dates FDA Received04/30/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/30/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTRIDGE MODEL-SFC-45, LOT#-UNK; FOAM TIP PLUNGER MODEL-FTP, LOT#-UNK; INJECTOR MODEL-MSI-PF, LOT#-UNK
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-