• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0706-XTW
Device Problem Incomplete Coaptation (2507)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2024
Event Type  Injury  
Event Description
It was reported that a mitraclip procedure was performed to treat functional mitral regurgitation (mr) with a grade of 4 and thin posterior leaflet.A mitraclip xtw was inserted and deployed on the mitral valve.However, after deployment, the clip detached from the posterior leaflet and remained attached to the anterior leaflet (single leaflet device attachment/slda).To stabilize the slda clip, an additional clip was implanted, reducing the mr to a grade of 2.Difficulties visualizing the clip during the procedure occurred.There was no clinically significant delay in the procedure.
 
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history did not indicate a lot-specific product issue.All available information was investigated, and the reported slda associated with the clip detaching from the posterior leaflet appears to be related to patient conditions (thin posterior leaflet).Image resolution poor is related to procedural circumstances as imaging was reported to be difficult.Unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key19119171
MDR Text Key340304675
Report Number2135147-2024-01719
Device Sequence Number1
Product Code NKM
UDI-Device Identifier05415067037381
UDI-Public(01)05415067037381(17)240718(10)30720R1107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCDS0706-XTW
Device Lot Number30720R1107
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/22/2024
Initial Date FDA Received04/16/2024
Supplement Dates Manufacturer Received05/07/2024
Supplement Dates FDA Received05/20/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
Patient Outcome(s) Required Intervention;
Patient Age59 YR
Patient SexFemale
-
-