BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS (ST); DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955596
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2024
Event Type  malfunction  
Manufacturer Narrative
Prismaflex st150 set c has been temporarily approved for use in the us under emergency use authorization eua (b)(4) to deliver crrt to treat patients in an acute care environment during the covid-19 pandemic.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an external fluid leak was observed from a ¿crack¿ in the access line at the filter connector of a prismaflex st150 set during priming.There was no patient involvement.No additional information was provided.
 
Manufacturer Narrative
The actual device was not available; however, photographs and a video of the sample were provided for evaluation.Visual inspection of the photos and video showed fluid leakage from the access blood header port.The specific defect that allowed the leak was not visible.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the condition was not determined as no further or actual sample testing could be performed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SETS (ST)
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19119182
MDR Text Key340304771
Report Number8010182-2024-00174
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414124403
UDI-Public(01)07332414124403
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955596
Device Lot Number22L0020
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/29/2024
Initial Date FDA Received04/16/2024
Supplement Dates Manufacturer Received05/16/2024
Supplement Dates FDA Received05/23/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-