|
Model Number URF-P7 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/08/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device evaluation is ongoing.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis results are pending.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
|
|
Event Description
|
It was reported that, during reprocessing, the uretero-reno fiberscope tested positive for 4 colony forming units (cfus) of moraxella and brevundimonas species.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacture's (lm) review of the customer cleaning disinfection and sterilization (cds) processes, results of third-party testing, the device evaluation and the lms final investigation.The lm reviewed the customer provided the cds processes where no obvious deviations from instructions for use (ifu) were identified.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: on (b)(6) 2024 sampling from: all channels cfu: <1cfu bacterial identification: n/a the device was evaluated where no abnormalities were found that could have led to the reported event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the ifu before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|