• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL (CVD-MINNETONKA) NEUROTHERM¿ GENERATOR TO DISPOSABLE US ELECTRODE, ADAPTER CABLE; CABLE, ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL (CVD-MINNETONKA) NEUROTHERM¿ GENERATOR TO DISPOSABLE US ELECTRODE, ADAPTER CABLE; CABLE, ELECTRODE Back to Search Results
Model Number DAC-NT
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/26/2024
Event Type  Injury  
Manufacturer Narrative
E1: reporter phone number: (b)(6).Unique device identifier (udi #): the udi is unknown because the lot number was not provided.During processing of this incident, attempts were made to obtain complete device information.Further information was requested but not received.
 
Event Description
Related manufacturer reference number: 3005334138-2024-00142 it was reported that during a procedure a message was received stating the electrode has been changed.As such, the treatment was stopped, and the patient was changed to medication treatment.Reportedly, there were no adverse consequence to the patient.A replacement adapter and same needle were used but the issue persisted.However, when troubleshooting with a representative¿s adapter cable, the issue was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROTHERM¿ GENERATOR TO DISPOSABLE US ELECTRODE, ADAPTER CABLE
Type of Device
CABLE, ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL (CVD-MINNETONKA)
14901 deveau place
minnetonka MN 55345 2126
Manufacturer (Section G)
ST. JUDE MEDICAL (CVD-MINNETONKA)
14901 deveau place
minnetonka MN 55345 2126
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key19121554
MDR Text Key340339171
Report Number2182269-2024-00009
Device Sequence Number1
Product Code IKD
UDI-Device Identifier05415067028884
UDI-Public(01)05415067028884(10)8656761
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K083528
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDAC-NT
Device Lot Number8656761
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/26/2024
Initial Date FDA Received04/16/2024
Supplement Dates Manufacturer Received04/19/2024
Supplement Dates FDA Received05/16/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/09/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-