(b)(4).Date sent: 4/16/2024.D4: batch # x7048h.Investigation summary.The product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned with the shaft bent.No functional testing could be performed due to the returned condition of the device.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, no damages was found on the jaws.Fourteen(14) clips were found inside clip track.Possible causes for the damage found may be inadvertent pressure placed on the device shaft through bending, pressing against the trocar, other devices on the back table, using the device as a retractor or moving heavy tissue with the device shaft.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
|