This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Additionally, to provide correction to the initial with information inadvertently left out (b5, d10, e2, e3, and g2) and to provide a correction to the initial (d9).The device was evaluated by olympus and confirmed the result of the reported event.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it is likely that the suggested event occurred due to external stress, the user could have applied force toward incorrect direction.Additionally, olympus measured fluid pressure of the reprocessor, though it was within the standard value, the pressure was higher than normal.Therefore, we presumed that load applied to the connecting tube was high, which caused connecting tube to be broken.Furthermore, olympus measured fluid pressure of the reprocessor, though it was within the standard value, the pressure was higher than normal.Therefore, we presumed that load applied to the connecting tube was high, which caused connecting tube to be broken.The root cause of the suggested event could not be identified.The event can be detected/prevented by following the instructions for use which state: preparation and inspection: 1.) visually inspect the connecting tube to ensure that there are no cracks, breaks, rips, scratches, attached debris, or other irregularities.2.) move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken.Additionally, olympus recommended the need to maintain temperatures ideally between 65-70 degrees to prevent further complications, particularly during warmer seasons.Olympus will continue to monitor field performance for this device.
|