Additional information: d4, d9, h3, h4, h6, h10.H4: the lot was manufactured between september 13, 2023 - september 15, 2023.H10: the actual device was received for evaluation.Visual inspection on the sample showed the bladder contained fluid and droplets of water located on the interior wall of the bottle.No evidence of rupture or leak was observed from the bladder or other components inside the bottle.A leak (rupture) test was performed, no evidence of leak or rupture was observed from the bladder.The bladder was found to be in normal condition.The reported condition of a bladder rupture was not verified; however, water droplets inside the bottle were observed which suggested a leak.The cause of the leak could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|