MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. CONNECTING TUBE; ENDOSCOPE REPROCESSOR ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. CONNECTING TUBE; ENDOSCOPE REPROCESSOR ACCESSORIES Back to Search Results
Model Number MAJ-2110
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  malfunction  
Manufacturer Narrative
To date, the customer have not returned the device for evaluation.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
It was reported the connecting tube had broken connectors.The issue occurred during unknown event.There were no reports of patient harm.
 
Manufacturer Narrative
B5, e2, e3, & g2: additional information has been obtained and provided.This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The dhr was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The case is not due to design.Based on the results of the legal manufacturer's investigation a definitive root cause of the reported event could not be identified.However, it¿s likely the cause is related to either of the following: -external stress was possibly applied to lock lever of the connecting tube, which led to breakage of the tube.Subsequently, the tube was unable to be connected.The user may have applied stress toward the wrong direction which caused the external stress.-the water pressure of the feeding fluid was within the standard value but higher than normal value.The connecting tube may have been broken by stress towards the tube.-the temperature of the supplied water was within the standard value but higher than the normal temperature.The high temperature of the water may have accelerated deterioration of the connecting tube, which led to the breakage.The suggested event is detectable by handling the device in accordance with the following instructions for use: 9 preparation and inspection.1 visually inspect the connecting tube to ensure that there are no cracks, breaks, rips, scratches, attached debris, or other irregularities.2 move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken.Olympus will continue to monitor field performance for this device.
 
Event Description
Per additional information obtained, the customer replaced the adapter and the product will not be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTING TUBE
Type of Device
ENDOSCOPE REPROCESSOR ACCESSORIES
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19124618
MDR Text Key341407330
Report Number9610595-2024-08026
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170404054
UDI-Public04953170404054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190969
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 05/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAJ-2110
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/19/2024
Initial Date FDA Received04/17/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/30/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-