The complaint was escalated for technical investigation.A product support engineer (pse) reviewed the clinical audit log, and the log shows that on (b)(6) 2024, at 08:06:05, an asystole alarm was generated.The asystole alarm was then acknowledged by the user at the central at 08:08:51.Then, the system generated several reminders every three minutes, as the condition was still present.At 08:26:23, the asystole alarm ended.Based on the information available, there was no product malfunction found with the device, as the device alarmed for asystole and was acknowledged by the user, and the device reminded for this red alarm.The reported problem was not confirmed.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|