Model Number 865240 |
Device Problem
No Audible Prompt/Feedback (2282)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/08/2024 |
Event Type
malfunction
|
Event Description
|
It was reported the intellivue patient monitor mx800 displays a speaker malfunction inop error message.It is unknown if there was still sound present to the device.A loss of audio cannot be ruled out based on information currently available.The device was in use on a patient.There was no report of patient or user harm.
|
|
Manufacturer Narrative
|
A follow up report will be submitted after the device has been received by philips for evaluation.(b)(6).
|
|
Manufacturer Narrative
|
A philips field service engineer (fse) went to the customer¿s site and tested the device.The fse confirmed the reported problem that the device displays a loudspeaker failure message.The fse replaced the speaker assembly to resolve the issue and confirmed the device was working as intended.A good faith effort (gfe) was conducted to clarify if the speaker produced audible sound; however, no response was received.
|
|
Event Description
|
It was reported the intellivue patient monitor mx800 displays a speaker malfunction inop error message.It is unknown if there was still sound present to the device.The device was in use on a patient.There was no report of patient or user harm.
|
|