MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM X; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
|
Back to Search Results |
|
Model Number 97800 |
Device Problem
High impedance (1291)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/09/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Information was received from a manufacturer representative (rep) regarding a patient with an implanted neurostimulator (ins) for urinary dysfunction/sacral nerve stim and gastrointestinal/pelvic floor therapy.The rep reported that it was reported that they first had impedances on two electrodes, then 3, then 4 when re-testing.Took components apart and dried them.They reinserted lead and tested again.Then had 4 impedances again. the lead and generator were taken apart, connected multiple times and tested multiple times.And continued to have an impedance.The surgeon requested to open a new battery.Upon testing that battery,.There was an impedance on the case only.Issue was resolved.
|
|
Manufacturer Narrative
|
Product analysis #706451894:analysis informationpli# 10 product id# 97800 below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection, output and telemetry testing, and functional testing.The implantable neurostimulator (ins) passed functional testing.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Additional information was received and it stated that there was a possible generator issue and failed impedance.
|
|