• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Decrease in Pressure (1490)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2024
Event Type  malfunction  
Event Description
Philips received a complaint from the customer, reporting that the v60 ventilator had an epap (expiratory positive airway pressure) reading that was lower than the set value.The device was in clinical use when the issue occurred.No medical intervention or delay in therapy was reported.It was reported that the patient was able to breathe spontaneously; there was no patient or user injury reported.The display pressure was observed as 6-7 when the device was set to 8.It was noted that the value which was set was displayed around once in 4 times of breath.The investigation is ongoing.
 
Manufacturer Narrative
H10 g - contact office phone number: (b)(6).E1 - reporter phone number - (b)(6).
 
Manufacturer Narrative
A service engineer (se) evaluated the device and reported that the alleged malfunction was not confirmed.There was no abnormality found after performing a functional test for an investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19125617
MDR Text Key340685620
Report Number2518422-2024-20305
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838033832
UDI-Public00884838033832
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1076709
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/31/2024
Initial Date FDA Received04/17/2024
Date Device Manufactured06/13/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-