This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.The fiber was broken directly at the end of the strain relief point, with strain relief side of the break having clear signs of burning with discoloration and melted plastic.Mechanical testing (rfid) found the device had worked as expected.Based on the results of the investigation, the event was primarily linked to procedural factors (device handling); however, a definitive root cause could not be determined.The event can be detected/prevented by following the instructions for use: soltive¿ superpulsed laser fiber single-use it was found that the device handling section states "do not bend or coil the soltive laser fibers beyond the recommended minimum bend radius (table 2); doing so may result in light leakage or fiber breakage that could cause personal injury if the laser is fired (refer to the laser system manual)." and device preparation section states "if any damage is observed such as breaks, kinks or damaged components, do not use the fiber, retain the device for manufacturer notification and use a replacement fiber".Olympus will continue to monitor field performance for this device.
|