Model Number OER-PRO |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/10/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported the oer-pro buttons were not working.The process could not proceed even when the confirm button was pressed after replacing the water filter.The issue occurred during reprocessing.There were no reports of patient harm.
|
|
Event Description
|
The issues has been resolved with the unit, error message code e51 was displayed.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over five (5) years since the subject device was manufactured.The device was not returned to olympus for inspection, therefore the customer's reportable malfunction could not be confirmed.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the process does not proceed even if the confirm button is pressed after replacing the water filter occurred due to the integrated chip in main control panel/sub control panel had failure by accumulated stress by use of the equipment or accidental factor to cause.Olympus will continue to monitor field performance for this device.
|
|