Manufacturer Narrative
A philips remote service engineer (rse) spoke with the customer and confirmed the reported problem "multiple surveillance and overviews are down".The rse logged in to the primary server and found that multiple surveillance and overviews were offline which were in reconnecting and service mode.A rse checked the switches under device status and there were six switches including two distribution switches and one coreb router that were offline.The rse asked biomed to restart the core b and after the restart some of the systems started working.After all the switches and the router were connected, the system started working but some of them were still in the service mode.The rse started placing the service mode systems into monitoring mode and that resolved the rest of the issue.The cause of the reported problem is network issue.The reported problem was confirmed.The device remains in use.
 
Event Description
The customer reported that multiple surveillance and overviews are offline.The device was in use on a patient at the time of the event.There was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT INFORMATION CENTER IX
Type of Device
PATIENT INFORMATION CENTER IX
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key19130403
MDR Text Key340563578
Report Number1218950-2024-00269
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838093041
UDI-Public00884838093041
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183387
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866389
Device Catalogue Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/25/2024
Initial Date FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/24/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse