• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH SRD S5 SINGLE ROLLER PUMP 150; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH SRD S5 SINGLE ROLLER PUMP 150; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/21/2024
Event Type  malfunction  
Event Description
Livanova deutschland received a report that , during installation, the s5 single roller pump 150 pump stopped and displayed alarm with yellow triangle shape including exclamation mark.No other activity was possible when the alarm was displayed.At the time of the first occurrence of the problem, the message 'motor controller failed pump 4' was displayed and, afterward, the message was not longer displayed.Additional tests found the problem recurs intermittently regardless of the operating state.There was no patient involvement.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 roller pump.The incident occurred in south korea.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
Through follow up, livanova was informed the problem with the roller pump was a problem during the installation process and the pump was not in use.As a result, the pump was not installed in the hospital and the pump was returned to the distributor.The event has been re-assessed as not reportable since issue was related to an out of box failure and would have been detected anyway prior device use on a patient through the functional tests to be performed at installation phase.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SRD S5 SINGLE ROLLER PUMP 150
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key19131987
MDR Text Key341132420
Report Number9611109-2024-00184
Device Sequence Number1
Product Code DWB
UDI-Device Identifier04033817900382
UDI-Public(01)04033817900382(11)221007
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/22/2024
Initial Date FDA Received04/18/2024
Supplement Dates Manufacturer Received04/22/2024
Supplement Dates FDA Received05/22/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/07/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-