The electronic log file was available for investigation.The case in question could be reconstructed by means of the information stored therein.During the case deviations between inspiratory and expiratory flow were detected, indicating a significant leakage in the patient circuit, leading to a fresh gas deficit in consequence.The log entries do not indicate a device failure.The device detected the leak and reacted as specified for such situation as it posted several alarms (e.G.Apnea, apnea co2, pressure high and fg low or leak).
|