• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH S5 ROLLER PUMP 150 (S5 RP150); CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH S5 ROLLER PUMP 150 (S5 RP150); CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Pumping Stopped (1503)
Patient Problem Low Blood Pressure/ Hypotension (1914)
Event Date 03/16/2024
Event Type  malfunction  
Manufacturer Narrative
A.1.-a.5.Patient information were not provided.H10: livanova deutschland manufactures the s5 roller pump 150.The incident occurred in france.Livanova intiated and investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received report that a s5 roller pump 150 pump made an overspeed alarm durin procedure.Specifically, after 5 minutes of cec, the arterial pump stopped for the first time and restarted by itself.Then 30 seconds later the arterial pump stopped again without the possibility of restarting it and displayed alarm: "acceleration error".Medical team, rescued the procedure with emergency hand crank, changed the arterial pump out and resumed the cec.The patient experience hypotension for approximately 1 minute.There is not report of any patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP 150 (S5 RP150)
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key19133049
MDR Text Key340683230
Report Number9611109-2024-00186
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/19/2024
Initial Date FDA Received04/18/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/04/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-