It was reported that noise oversensing and pacing inhibition was noted on this right ventricular (rv) lead.The patient experienced asystole and was hospitalized for monitoring.Technical services (ts) reviewed device data and provided troubleshooting recommendations.The patient was noted to have chronic atrial fibrillation (af), and xray confirmed that the rv tip and shocking coil was located higher than recommended.Provocation maneuvers were performed, and the noise was reproduced on the rv lead.This rv lead is expected to be replaced and returned to boston scientific for analysis.No adverse patient effects were reported.Additional information was received.Loss of capture was also noted on the rv lead.This rv lead was successfully replaced and is expected to be returned for analysis.No additional adverse patient effects were reported.
|