It was reported that the patient presented for an implant procedure.During the procedure, it was noted that the right atrial lead (ra) exhibited a high capture threshold and low atrial sensing issue.While attempting to change positions, the helix of the ra lead exhibited difficulty in retracting.Additionally, the right ventricular lead (rv) was found to have a helix retraction issue and a crystalline object fell out of the lead.The ra and rv leads were not used and replaced during the procedure.The patient experienced no consequences.
|
The reported events were failure to retract helix, high capture thresholds and low sensing.As received, a complete lead was returned in one piece with the helix found bent due to procedural damage and clogged with blood.The reported event of helix mechanism issue was confirmed.X-ray inspection found over-torque of the inner coil at the connector region consistent with procedural damage.X-ray examination of the helix mechanism found the helix bent consistent with procedural damage.The cause of the reported event of failure to retract helix was isolated to helix bent and over-torqued of the inner coil.The reported events of high capture thresholds and low sensing were not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspections of the lead did not find any anomalies except for procedural damage.
|