• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE X3

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE X3 Back to Search Results
Model Number 867030
Device Problem No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2024
Event Type  malfunction  
Manufacturer Narrative
A philips field service engineer (fse) visited onsite and confirmed the reported issue that spo2 is reading and then readings not showing on monitor.Customer was using nellcor disposable sensor.A fse tested x3 spo2 in empty room using pronk simulator and it was noted that signal and values were stable until various lights were turned on in the room.Fse determines that sensors are sensitivity to light interference and demonstrated this to customer nurse manager and biomed.The reported problem was confirmed.The x3 was found to be functioning correctly and the nellcor disposable sensor was at fault.
 
Event Description
Customer reported the intellivue x3 had the following issue: ¿spo2 is reading and then readings not showing on monitor".The device was in use at the time of the event.No patient or user harm or injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE X3
Type of Device
INTELLIVUE X3
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key19133776
MDR Text Key340493428
Report Number9610816-2024-00206
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838082588
UDI-Public00884838082588
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K171801
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number867030
Device Catalogue Number867030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/04/2024
Initial Date FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-