• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS MX40 1.4 GHZ SMART HOPPING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS MX40 1.4 GHZ SMART HOPPING Back to Search Results
Model Number 865350
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/20/2024
Event Type  malfunction  
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Event Description
The customer indicated that the device did not alarm as intended.It is unknown if the device was in use at the time of the event.There was no adverse event reported.
 
Manufacturer Narrative
Source notes indicated "device alarm/alert as intended? : no", but the gfe response from the biomedical equipment technician stated that there was no fault with the device and the asset was being returned from bench repair.Bench repair traveler id # (b)(4) opened on 11mar2024 indicated that the device was repaired on 12mar2024 and returned to the customer.The gfe response indicated that there is no fault of the device, the device is functioning as designed and is back in service at the customer site.The reported problem was not confirmed.The biomedical equipment technician indicated that there is no fault of the device, the device is functioning as designed and is back in service at the customer site.Correction to a1-a6.Device use is updated to outside of use and questions were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MX40 1.4 GHZ SMART HOPPING
Type of Device
MX40 1.4 GHZ SMART HOPPING
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key19134308
MDR Text Key340519680
Report Number1218950-2024-00271
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838082236
UDI-Public00884838082236
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865350
Device Catalogue Number865350
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/20/2024
Initial Date FDA Received04/18/2024
Supplement Dates Manufacturer Received04/09/2024
Supplement Dates FDA Received04/23/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-