• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB EASY-LOAD LENS DELIVERY SYSTEM; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB EASY-LOAD LENS DELIVERY SYSTEM; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Model Number EZ-28V
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  Injury  
Manufacturer Narrative
Although requested, the device was not returned for evaluation and additional information regarding the event was not provided.A review of the device history record (dhr) did not identify any anomalies or nonconformities that could be related to this event.The lot history, trend analysis, risk analysis and directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the available information, the root cause could not be conclusively determined; however, user related factors (such as loading or handling techniques) and/or procedural factors (such as lens and inserter interaction) might have contributed to the event.
 
Event Description
Reportedly, during implantation of an intraocular lens (iol) into the eye, it was noticed that a haptic was bent.The incision was enlarged, the iol was removed, and a backup lens was implanted with no adverse event noted.Additional information was requested, but not received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASY-LOAD LENS DELIVERY SYSTEM
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
BAUSCH + LOMB
1400 north goodman street
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
21 north park place blvd.
clearwater FL 33759
Manufacturer Contact
shayan habibi
21 n park pl blvd
clearwater, FL 33759
7277246600
MDR Report Key19134531
MDR Text Key340501600
Report Number0001313525-2024-00089
Device Sequence Number1
Product Code MSS
UDI-Device Identifier00757770520022
UDI-Public(01)00757770520022(17)240620(10)H2317001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberEZ-28V
Device Lot NumberH2317001
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SILICONE IOL
Patient Outcome(s) Required Intervention;
Patient Age82 YR
Patient SexFemale
-
-