Model Number L111
Device Problems Premature Discharge of Battery (1057); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/25/2024
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the battery of this implantable pacemaker device was suspected to be depleting prematurely.This device was implanted for approximately a year now and the battery shows five years remaining.Boston scientific technical services (ts) noted that there was a slight increase in power consumption and the patient was in chronic atrial fibrillation (af).Ts also discussed the impact of high-rate atrial sensing on battery longevity, and that device is of standard longevity.Device reprogramming was recommended, and the patient will continue to be monitored.This device remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESSENTIO MRI DR
Type of Device
IMPLANTABLE PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19134559
MDR Text Key340829157
Report Number2124215-2024-23619
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526558962
UDI-Public00802526558962
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberL111
Device Catalogue NumberL111
Device Lot Number665164
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/26/2024
Initial Date FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age86 YR
Patient SexFemale
-
-