• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CONTOUR VL; STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CONTOUR VL; STENT, URETERAL Back to Search Results
Model Number M0061801560
Device Problems Device-Device Incompatibility (2919); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2024
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that a contour vl ureteral stent was used during a stent insertion procedure performed on (b)(6) 2024.During the procedure, when the physician attempted to remove the sensor wire for stent placement, the sensor wire was stuck onto the stent.The stent became shredded, and the wire uncoiled itself.The guidewire and stent were removed from the patient without a problem.Another stent and guidewire completed the procedure.There were no patient complications reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
 
Manufacturer Narrative
Block h6: imdrf device code a0414 captures the reportable event of stent torn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR VL
Type of Device
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key19134820
MDR Text Key340832649
Report Number2124215-2024-22571
Device Sequence Number1
Product Code FAD
UDI-Device Identifier08714729067986
UDI-Public08714729067986
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K974541
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM0061801560
Device Catalogue Number180-156
Device Lot Number0032630059
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/25/2024
Initial Date FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-