Model Number 1948/52
Device Problems Intermittent Capture (1080); Fracture (1260); High impedance (1291); Over-Sensing (1438)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2024
Event Type  Injury  
Event Description
It was reported during in clinic follow up that the right ventricular lead exhibited oversensing noise, high pacing impedance, and intermittent capture.Imaging revealed that the lead had fractured.The lead was capped on (b)(6) 2024 and successfully replaced.The patient was stable and asymptomatic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOFLEX OPTIM LEAD
Type of Device
Permanent defibrillator electrodes
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19135957
MDR Text Key340517803
Report Number2017865-2024-39048
Device Sequence Number1
Product Code NVY
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P960030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2014
Device Model Number1948/52
Device Lot Number3414631
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/01/2024
Initial Date FDA Received04/18/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/27/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ACCENT
Patient Age87 YR
Patient SexMale
-
-