• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/15/2024
Event Type  malfunction  
Manufacturer Narrative
G1 contact office phone: (b)(6).
 
Event Description
Philips received a complaint on the v60 ventilator, indicating that there was a proximal pressure sensor autozero failed alarm.The device was reported to be in use at the time of the reported problem.No patient or user harm reported.The remote service engineer (rse) informed the customer to perform the following troubleshooting steps in this order: verify the pin alignment between the solenoids and data acquisition (da) printed circuit board assembly (pcba), replace the 3rd and 4th proximal autozero solenoids, and then replace the da pcba.This investigation is ongoing.
 
Manufacturer Narrative
H10: in a good faith effort (gfe) response from the customer received on 01may2024, it was stated that the customer had made the decision to pull the ventilator which experienced the autozero alarm from service.The customer was unsure if the repair would be completed in the future, if the device would be traded in, or discarded.No further work was performed by philips to resolve the issue.The customer did not provide the patient information in the gfe response received on 01may2024.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19137096
MDR Text Key340869215
Report Number2518422-2024-20770
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/15/2024
Initial Date FDA Received04/18/2024
Supplement Dates Manufacturer Received05/01/2024
Supplement Dates FDA Received05/02/2024
Date Device Manufactured09/10/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-