It was reported, the endoscope reprocessor's a2 connector spring came out as the middle pin and spring separated from the rest of the connector.Olympus field service engineer went onsite and verified that the pin was missing from the a2 connector.Connector was able to be replaced, and the machine was able to then function properly.The issue occurred during reprocessing, and there were no reports of patient harm.
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information added to field d8, h3, h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was not returned to olympus for inspection, and the customer's complaint of a2 connector had spring come out was confirmed.Based on the results of the investigation, it is likely the loosen of a2 connector in reprocessing basin made middle pin and spring inside the connector came off.As for the cause of the loosen a2 connector, stress to the connector toward loosening direction may have been accumulated, making the connector came off.However, a definite root cause that led to the malfunction could not be determined.The event can be detected and prevented by following the instructions for use which state: chapter 3.Inspection before use: 5.6 inspecting the connectors: check the following for each connector.The connector should be fixed firmly.The o-rings should be free of irregularities such as cracks, tears, or dents.If any irregularity is found, do not use the reprocessor and contact olympus.[warning] do not use the equipment if any connector seems to be damaged or defective.Using the equipment when an irregularity has been detected may interfere with reprocessing.Furthermore, fluid leakage may damage peripheral devices or facilities near the equipment.Olympus will continue to monitor field performance for this device.
|