MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM X; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
|
Back to Search Results |
|
Model Number 97800 |
Device Problems
Communication or Transmission Problem (2896); Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/28/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Information was received from an healthcare provider (hcp) regarding a patient who was being implanted with an implantable neurostimulator (ins) for urinary dysfunction (incontinence, urgency, and/or frequency).It was reported that the hcp was unable to connect to check impedance after several tries with handset.Blood in port is the suspected issue.New ins was used and it connected right away with impedance in range.
|
|
Manufacturer Narrative
|
H3: analysis of the ins (sn: (b)(6) ) revealed that the implantable neurostimulator (ins) passed functional testing.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|