This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the events were confirmed, and the device did not meet the standard specifications and function.The probable cause is that electrical load or stress had been applied to the electrical board in the image sensor due to energization.This was mounted on the image sensor unit embedded in the distal end of the endoscope.It was likely that static electricity had been inadvertently applied, or overcurrent had flown due to connection/disconnection while the system was powered on.Then, the electrical connecting status temporarily became poor, which exercised a negative impact to the image signal output to make it unstable.As a result, the image sensor unit became faulty (irregular), leading to the image irregularity.Moreover, overcurrent may have been applied due to connection/disconnection while the system was powered on, overcurrent from other devices or electrical wiring, or adhesion of dust and moisture to the electrical contacts on the system and those in the video connector.The instruction manual identifies the following related verbiage which could have detected the phenomenon: instructions, operation manual, chapter 3 "preparation and inspection", section 3.8 "inspection of the endoscopic system".Olympus will continue to monitor field performance for this device.
|