Model Number 470183-16 |
Device Problem
Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/19/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the permanent cautery hook instrument had a cable issue that was identified prior to starting a procedure.No further specifics were provided.There was no allegation of patient injury since the instrument was not used during the procedure as planned.
|
|
Manufacturer Narrative
|
The permanent cautery hook instrument has been received for failure analysis testing, but the evaluation is still in progress.
|
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The permanent cautery hook instrument was analyzed, and the findings did not replicate or confirm the customer reported event.Failure analysis found the primary finding of expected condition to be related to the customer reported complaint.Visual inspection displayed no signs of physical damage.The instrument was returned with a reported complaint that does not affect functionality and is a part of the instruments design.The instrument was reported as missing an input disk however with this instrument does not require an extra input disk in the reported area.The instrument is in its expected condition.No product issue was identified, and the product is considered conforming in its current state.
|
|
Event Description
|
Refer to h10/h11 for follow-up information.
|
|