• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHAIR; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHAIR; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735249
Device Problems Use of Device Problem (1670); Mechanics Altered (2984)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/22/2023
Event Type  malfunction  
Manufacturer Narrative
H3 - evaluation of the returned reference frame determined that as reported, the weld has broken on the base of the starburst arm leaving the frame loose.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
Medtronic received information regarding a navigation system during a sacroiliac and thoracolumbar procedure.It was reported that during a case, the medtronic representative (rep) discovered that the reference frame was wobbly and looked as if a weld was broken on it.The reference frame was loose.There was less than an hour delay in procedure.There was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHAIR
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key19140948
MDR Text Key341430980
Report Number1723170-2024-01113
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number9735249
Device Catalogue Number9735249
Device Lot Number210914
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2024
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age71 YR
Patient SexFemale
Patient Weight51 KG
-
-