• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; NO MATCH Back to Search Results
Model Number CDHFA500Q
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pocket Erosion (2013)
Event Type  Injury  
Event Description
It was reported, that the patient presented in the clinic.With device and leads eroded from the pocket.The entire implantable cardioverter defibrillator was visible from the opened incision site of the implanted device pocket.While all the right ventricular, left ventricular and atrial leads were visible in the proximal part of the leads.The physician explanted the entire system implantable cardioverter-defibrillator, right ventricular lead, left ventricular lead and atrial lead, due to erosion.The patient was in stable condition throughout the procedure.
 
Manufacturer Narrative
The reported event was pocket erosion.Analysis found that the device was received with battery voltage above elective replacement indicator (eri) level.Interrogation results were normal, visual screen was acceptable, and preliminary test results were acceptable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19140995
MDR Text Key340561976
Report Number2017865-2024-39097
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public(01)05415067032010(10)S000091614(17)250731
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberCDHFA500Q
Device Lot NumberS000091614
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/18/2024
Supplement Dates Manufacturer Received04/30/2024
Supplement Dates FDA Received05/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age67 YR
Patient SexMale
Patient Weight141 KG
-
-